Anion exchange membrane test protocol validation

Publication Type

Journal Article

Date Published

04/02/2025

Authors

DOI

Abstract

This study presents the validation of protocols for measuring ion exchange capacity (IEC) and alkaline stability of anion exchange membranes (AEMs) for low-temperature water electrolysis. While protocols are often tested within individual laboratories, their results across multiple laboratories with varying equipment, environmental conditions, and personnel qualification remain unverified. The validation involved Los Alamos National Laboratory (LANL), National Renewable Energy Laboratory (NREL), and University of Oregon (UO) using the same commercially available AEM to assess reproducibility and reliability of the protocols under diverse conditions. For the IEC protocol, results across laboratories were consistent within ±10% of the NMR-determined reference value. The alkaline stability protocol could pose greater challenges due to factors such as variations in sample collection timing, preservation methods, and analytical techniques, but consistent test results for percentage IEC loss were demonstrated across institutions. These results highlight the reliability and applicability of the protocols, emphasizing the importance of validation to ensure consistency in diverse research environments.

Journal

Frontiers in Energy Research

Volume

13

Year of Publication

2025

URL

ISSN

2296-598X

Organization

Research Areas